Changing Your Pod 3
3 Changing Your Pod
Warning:
Check the infusion site after insertion to ensure that the cannula
was properly inserted. If the cannula is not properly inserted, hyperglycemia
may result.
4. If there is a problem with the cannula, press
No
and
follow the on-screen instructions for deactivating
your Pod. Then restart the process with a new Pod.
5. Press
Yes
if the cannula was properly inserted. The
PDM shows the
Status
screen.
6. After 1.5 hours, the PDM prompts you to check your blood glucose level and
to recheck the infusion site.
Warning:
NEVER inject insulin (or anything else) into the fill port while
the Pod is on your body. Doing so may result in unintended or interrupted
insulin delivery.
Avoid Infusion Site Infections
Check the infusion site at least once a day:
• Be aware of signs of infection, including pain, swelling, redness, discharge, or
heat at the site. If you suspect an infection, immediately remove the Pod and
apply a new one in a different location. Then call your healthcare provider.
• If you observe any problems with the Pod, replace it with a new Pod.
Summary of Contents for UST400
Page 27: ...12 Thispageintentionallyleftblank ...
Page 37: ...22 Thispageintentionallyleftblank ...
Page 61: ...46 Thispageintentionallyleftblank ...
Page 69: ...54 Thispageintentionallyleftblank ...
Page 99: ...84 Thispageintentionallyleftblank ...
Page 109: ...94 Thispageintentionallyleftblank ...
Page 147: ...132 Thispageintentionallyleftblank ...
Page 187: ...172 Thispageintentionallyleftblank ...
Page 201: ...My Notes ...