167
Appendix
Appendix
Electromagnetic Immunity
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast, and TV broadcast, cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
equipment is used exceeds the applicable RF compliance level above, the equipment
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the equipment.
Recommended Separation Distances Between Portable and
Mobile RF Communications Equipment and the System
You can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the System, as recommended below, according to the
maximum output power of the communications equipment.
Rated
maximum
output power
of transmitter,
in watts
Separation distance according to frequency of transmitter,
in meters
150 kHz to 80 MHz
d = 1.17 √P
80 MHz to 800 MHz
d = 0.35 √P
800 MHz to 2.5 GHz
d = 0.7 √P
0.01
0.12
0.035
0.070
0.1
0.37
0.11
0.22
1
1.17
0.35
0.7
10
3.70
1.11
2.21
100
11.7
3.5
7.0
For transmitters rated at a maximum output power not listed above, the recommended
separation distances in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Summary of Contents for UST400
Page 27: ...12 Thispageintentionallyleftblank ...
Page 37: ...22 Thispageintentionallyleftblank ...
Page 61: ...46 Thispageintentionallyleftblank ...
Page 69: ...54 Thispageintentionallyleftblank ...
Page 99: ...84 Thispageintentionallyleftblank ...
Page 109: ...94 Thispageintentionallyleftblank ...
Page 147: ...132 Thispageintentionallyleftblank ...
Page 187: ...172 Thispageintentionallyleftblank ...
Page 201: ...My Notes ...