155
Appendix
Summary of Settings and Options
The options for the various Omnipod® Insulin Management System settings are:
Time
12-hour or 24-hour clock
Date
MM/DD/YY
DD/MM/YY
MM.DD.YY
DD.MM.YY
YY-MM-DD
Maximum Basal Rate
0.05-30 U/hr. Default is 3.00 U/hr
Basal rate
U/hr. Range: 0.05 U/hr to Maximum Basal Rate in 0.05
U/hr increments.
Basal Programs
7
Basal rate segments
24 per program
Temp basal
%, U/hr, or Off. Default is Off
Duration: 30 min to 12 hrs in 30-min increments
Temp basal (set to %)
Range: 0 U/hr ("Off") to 95%MORE of current basal
rate in 5% increments. Cannot exceed Maximum Basal
Rate.
Temp basal (set to U/hr) Range: 0 0 U/hr ("Off") to Maximum Basal Rate
Temp basal presets
7
BG Goal Range for
blood glucose history
Lower and upper limits: 3.9 to 11.1 mmol/L in 0.1
mmol/L increments
BG reminder
On or Off. Default is Off.
Maximum of 4 active at one time
Reminder can occur between 30 min to 4 hrs after
bolus is started. Set in 15-minute increments.
Custom reminder
Maximum of 4. Set to Daily, One time only, Off
BG Meter sound
On or Off. Default is On.
Bolus Calculator
On or Off. Default is On.
Target BG value
8 time segments; 3.9 to 11.1 mmol/L in 0.1 mmol/L
increments
Correct Above threshold 8 time segments; Target BG to 11.1 mmol/L in 0.1
mmol/L increments
Minimum BG for Calcs 2.8 to 3.9 mmol/L in 0.1 mmol/L increments
Default is 3.9 mmol/L
Summary of Contents for UST400
Page 27: ...12 Thispageintentionallyleftblank ...
Page 37: ...22 Thispageintentionallyleftblank ...
Page 61: ...46 Thispageintentionallyleftblank ...
Page 69: ...54 Thispageintentionallyleftblank ...
Page 99: ...84 Thispageintentionallyleftblank ...
Page 109: ...94 Thispageintentionallyleftblank ...
Page 147: ...132 Thispageintentionallyleftblank ...
Page 187: ...172 Thispageintentionallyleftblank ...
Page 201: ...My Notes ...