105
Alarms, Notifications, and Other Messages 9
9 Alarms, Notifications, and Other Messages
Failure when deactivating a Pod
PDM message
What to do
Communication error.
Remove pod.
To retry deactivation, press
"Retry."
Otherwise, press
"Discard."
An attempt to deactivate the Pod fails.
Remove your Pod.
Press
Retry
to try deactivating the Pod again, or
press
Discard
to break the link between the PDM
and the unresponsive Pod.
Activate a new Pod.
Warning:
If you are unable to deactivate a Pod, it continues to pump insulin.
Be sure to remove the failed Pod before you activate a new Pod to prevent
hypoglycemia.
Special situation
A special situation exists if a bolus was being delivered when communication was
lost, because the PDM does not know how much of the bolus was delivered. This
causes the Bolus Calculator to be temporarily disabled because it cannot calculate
a valid IOB (insulin on board). The Bolus Calculator is re-enabled when no
insulin remains in your body from the old bolus. This length of time is determined
by your duration of insulin action.
PDM message
What to do
Due to a recent clock
reset or unconfirmed
bolus information, bolus
calculations cannot be
used until <time>.
The Bolus Calculator is disabled until the listed
time.
Press
OK
.
You can deliver a manually-calculated bolus if you
want.
Summary of Contents for UST400
Page 27: ...12 Thispageintentionallyleftblank ...
Page 37: ...22 Thispageintentionallyleftblank ...
Page 61: ...46 Thispageintentionallyleftblank ...
Page 69: ...54 Thispageintentionallyleftblank ...
Page 99: ...84 Thispageintentionallyleftblank ...
Page 109: ...94 Thispageintentionallyleftblank ...
Page 147: ...132 Thispageintentionallyleftblank ...
Page 187: ...172 Thispageintentionallyleftblank ...
Page 201: ...My Notes ...