6-1
Output Format
6
6.1
Functions as Talker
There are the following seven types of data output:
• Measured data output (ASCII code): TS0 + “Device Trigger (GET)” + FM0
• Measured data output (binary code): TS0 + “Device Trigger (GET)” + FM1
• Computed data output (ASCII code): TS0 + “Device Trigger (GET)” + FM2
• Computed data output (binary code): TS0 + “Device Trigger (GET)” + FM3
• Setting of data output in the operation mode: TS1 + “Device Trigger (GET)” + LF
• Unit and decimal point position data output: TS2 + “Device Trigger (GET)” + LF
• System configuration data output: TS5 + “Device Trigger (GET)” + CF
• A/D calibration data output: TS8 + “Device Trigger (GET)” + LF
• Setting of data output in the setup mode: TS9 + “Device Trigger (GET)” + LF
• Outputting of a file directory on a RAM disk: TS3 + [GET] + [MF0]
• Outputting of measured/computed data (ASCII code) on a RAM disk: TS3 + [GET] + MF1
• Outputting of the measured/computed data (binary code) on a RAM disk: TS3 + [GET] + MF2
• Outputting of channel on/off on a RAM disk: TS3 + [GET] + MF3
Measured data output (TS0/TS3)
After executing “GET,” be sure to output data using the FM or MF command. Execution of “GET”
alone without executing the FM or MF command does not output the data. After reading all the data
specified by the FM command, subsequent specification of the FM command without executing
“GET” enables the data within a scan to be output.
Because the MF command requests the output of data stored on the internal RAM disk, so a differential
in time occurs between the time when the data are measured and the time when [GET] is executed.
Measured data output (TS1, TS2, TS8 or TS9)
After transmitting the device trigger “GET ”be sure to output data using the LF command. Execution
of “GET” alone without executing the FM command does not output the data. After reading all the
data specified by the LF command, subsequent specification of another channel using the LF command
enables the data contents to be output. A/D calibration data output by the TS8 command can be
executed in the A/D calibration mode only.
System configuration output (TS5)
After transmitting the device trigger “GET ”be sure to output data using the CF command. Execution
of “GET” alone without executing the CF command does not output data.
Note
• When using an RS-232-C, RS-422-A or RS-485 interface, execute an ESCT command rather than a GET command.
• Do not transmit the FM, LF, CF or MF command before outputting data for specified channels.
• If an FM, LF, CF or MF command is received while data are being transmitted, transmission of the data is suspended
automatically.
• If the type of output data is changed using the TS command after the execution of “GET,” the changed contents are not
reflected without executing “GET” again. Execute “GET” again.
• After executing “GET,” the execution of “GET” again without outputting data using the FM, LF, CF or MF command or
without completing the data output sets new data to the buffer. Be careful because old data are lost.
Summary of Contents for DR130 SERIES
Page 39: ...4 9 Command Format 4 4 5 Input Range Parameter This page intentionally left blank ...
Page 40: ...4 10 4 5 Input Range Parameter This page intentionally left blank ...
Page 56: ...5 15 Commands 5 This page intentionally left blank 5 5 Other Settings ...
Page 71: ...6 9 Output Format 6 This page intentionally left blank ...