5-3
Commands
5
XQ
Sets the filter.
Mode
Setup mode
Setting
XQp1<terminator>
p1
ON
Filter on
OFF
Filter off
Example
Insert a filter in the input.
XQON
XV
Sets the measurement period.
Mode
Setup mode
Setting
XVp1<terminator>
p1
Measurement interval ( 2, 3, 4, 5, 6, 10, 12, 15,
20, 30, or 60)
Example
Set the measurement period to 10 seconds.
XV10
Comments
• The shortest measurement period that can be set
differs depending on the A/D integration time and
the filter on/off setting as follows.
Integration time: Equivalent to 50/60 Hz
Input module
Filter off
Filter on
10CH
2 s
3 s
20CH
2 s
4 s
30CH
2 s
4 s
Integration time: 100 ms
Input module
Filter off
Filter on
10CH
4 s
12 s
20CH
5 s
15 s
30CH
6 s
20 s
XI
Sets the A/D integration time.
Mode
Setup mode
Setting
XIp1, p2<terminator>
p1
Subunit (0 to 5)
p2Integration time (AUTO, 50 Hz, 60 Hz, 100
ms)
Example
Set the A/D integration time in subunit 0 to 50 Hz.
XI0, 50Hz
5.1 Setting the Input