2-9
Overview and Specifications of RS-232-C Interface
2
2.6
RS-232-C Interface Parameter Setting Procedure
Setting of the RS-232-C parameters must be carried out using the 3 dipswitches located next to the
module connector.
1
2
3
4
Parameter setting
switch 1
Data length
Baud rate
1
2
3
4
Parameter setting
switch 2
Not use
Parity
Stop bit
1
2
3
4
Parameter setting
switch 3
Not use
Handshake system
ON
OFF
OFF
OFF
ON
ON
Baud rate (Switch No.1 to 3 of parameter setting switch 1)
Baud rate
No.1
No.2
No.3
150
OFF
OFF
OFF
300
OFF
OFF
ON
600
OFF
ON
OFF
1200
OFF
ON
ON
2400
ON
OFF
OFF
4800
ON
OFF
ON
9600
ON
ON
OFF
←
Default
19200
ON
ON
ON
Data length (Switch No.4 of parameter setting switch 1)
Data length
No.4
7
OFF
8
ON
←
Default
Parity (Switch No.1 and 2 of parameter setting switch 2)
Parity
No.1
No.2
NONE
OFF
OFF
ODD
OFF
ON
EVEN
ON
OFF
←
Default
Stop bit (Switch No.3 of parameter setting switch 2)
Stop bit
No.4
1
OFF
←
Default
2ON
Handshake system (Switch No.1 to 3 of parameter setting switch 3)
Handshake system
No.1
No.2
No.3
No handshake
OFF
OFF
OFF
←
Default
XON-ER
OFF
OFF
ON
XON-RS
OFF
ON
OFF
CS-ER
OFF
ON
ON
CS-RS
ON
OFF
OFF
Summary of Contents for DR130 SERIES
Page 39: ...4 9 Command Format 4 4 5 Input Range Parameter This page intentionally left blank ...
Page 40: ...4 10 4 5 Input Range Parameter This page intentionally left blank ...
Page 56: ...5 15 Commands 5 This page intentionally left blank 5 5 Other Settings ...
Page 71: ...6 9 Output Format 6 This page intentionally left blank ...