5-19
Commands
5
RC
Initialize the set values.
Mode
Operation mode
Setting
RCp1<terminator>
p1
0 The set values are initialized.
Example
Initialize the operation mode parameters (measuring
range, unit, alarm, date & time, and moving average).
RC0
Comments
• Since it takes an indefinite time to execute the RC
command, return an ACK after processing. On the
controller side, execute the following processing
after receiving an ACK after transmitting the RC
command. The format of the ACK status is as shown
below.
E0Cr+Lf
The received command was normally
processed.
E1Cr+Lf
There is an error in the received command.
• After executing the command, the clock is initialized
to 96/01/01 00:00:00.
DS
Transfers the setting mode.
Mode
All modes
Setting
DSp1<terminator>
p1
0 Transfers to operation mode.
1 Transfers to setup mode.
2 Transfers to A/D calibration mode.
Example
Transfer the DR230/DR240 mode to the setup mode.
DS1
Comments
• Since it takes an indefinite time to execute the DS
command, return an ACK after processing. On the
controller side, execute the following processing
after receiving an ACK after transmitting the DS
command. The format of the ACK status is as shown
below.
E0Cr+Lf The received command was normally
processed.
E1Cr+Lf There is an error in the received
command.
5.7 Data Output Request
Command
TS
Selects the output data.
Mode
All modes (A/D calibration data output is limited to A/D
calibration mode only.)
Setting
TSp1<terminator>
p1
0 Measured data output
1 Setting data output
2 Unit data output
4 Report data output
5 System configuration data output
8 A/D calibration data output
9 Setting data output in setup mode
Comments
• The setting for p1=0, 1 or 2 is effective only in the
operation mode.
• The setting for p1=8 is effective only in the A/D
calibration mode.
• The setting “p1=3” is valid only for recorders with
a floppy disk drive function .
• Do not output data from the RAM disk using a TS3
command during computation.
• The setting, p1=4 is effective when the instrument
has the report function and one of hourly, daily and
monthly reports is ON.
FM
Selects the output format for measured/
computed data.
Mode
Operation mode
Setting
FMp1, p2, p3<terminator>
p1
0 Outputs measured data in ASCII format.
1 Outputs measured data in binary format.
2 Outputs computed data in ASCII format.
3 Outputs computed data in binary format.
p2First output channel (001 to 030)
First output computed channel (A01 to A30)
p3
Last output channel (001 to 030)
Last output computed channel (A01 to A30)
Comments
• You must specify data to be output with the TS
command and execute “GET” or “ESC T”before
sending the FM command.
• The command selects the output format of measured/
computed data, either ASCII or binary, and the
output channel.
• If no input channel is recognized by the DR130/
DR230/DR240 among the specified channels, a
syntax error occurs.
• The setting “p1=2” or “p1=3” is valid only for
recorders with the optional computing function or
floppy disk drive function.
• The optional computation channel numbers are
A01 to A30.
MF
Sets the output formats of the measured
data/computed data on a RAM disk.
Mode
Operation mode
Setting
MFp1, p2, p3, p4, p5, p6<terminator>
p1
Output contents
0
File directory
1
Measured/computed data output in ASCII
format.
2Measured/computed data output in binary
format
3
On/off information on channel on which
specified file exists
5.6 Control and Execution Command
Summary of Contents for DR130 SERIES
Page 39: ...4 9 Command Format 4 4 5 Input Range Parameter This page intentionally left blank ...
Page 40: ...4 10 4 5 Input Range Parameter This page intentionally left blank ...
Page 56: ...5 15 Commands 5 This page intentionally left blank 5 5 Other Settings ...
Page 71: ...6 9 Output Format 6 This page intentionally left blank ...