6-2
6.2
Measured/Computed Data Output Format (ASCII
code)
The data are output in the following format by receiving TS0 + “Device Trigger (GET)” + FM0/
FM2:
DATEYYMMDDCrLf
TIMEhhmmssCrLf
S1S2A1A1A2A2A3A3A4A4UUUUUUCCC,
±
DDDDDE - ECrLf
Each symbol denotes the following:
YY:
Year
MM:
Month
DD:
Day
hh:
Hour
mm:
Minute
ss:
Second
S1:
Data status 1
E ------------- Abnormal
N ------------ Normal
D ------------ Differential input
O ------------ Over
S ------------- Measuring range is “skip” or computation channel is “off”
S2:
Data status 2
Space ------- Interim data
E ------------- Last data
A1A1:
Alarm status (level 1)
A2A2:
Alarm status (level 2)
A3A3:
Alarm status (level 3)
A4A4:
Alarm status (level 4)
H
--------- Upper-limit alarm
L
---------- Lower-limit alarm
dH ----------- Upper-differential-limit alarm
dL ----------- Lower-differential-limit alarm
RH ---------- Increasing rate-of-change limit alarm
RL ----------- Decreasing rate-of-change limit alarm
UUUUUU: Unit
mV
--- mV
V
-- V
C
--
°
C
UUUUUU ------- Arbitrary
CCC:
Channel number
±
:
Data polarity (+, -)
DDDDD:
Data mantissa (8 characters for the computation channel)
±
99999 ----- Over data
+99999 ----- Abnormal data
E - E:
Data exponent
Note
• Data in the channel not connected in the system settings, including channel numbers, are not output.
• In a channel for data computation, the channel number CCC is indicated as follows:
DR130/DR230/DR240: A01 to A30
Summary of Contents for DR130 SERIES
Page 39: ...4 9 Command Format 4 4 5 Input Range Parameter This page intentionally left blank ...
Page 40: ...4 10 4 5 Input Range Parameter This page intentionally left blank ...
Page 56: ...5 15 Commands 5 This page intentionally left blank 5 5 Other Settings ...
Page 71: ...6 9 Output Format 6 This page intentionally left blank ...