5-1
Commands
5
5.1 Setting the Input
SR
Sets the range.
Mode
Operation mode
Setting the range to SKIP
Setting
SRp1, p2<terminator>
p1
Channel number (001 to 030)
p2SKIP
Example
Skip channel 01 of slot 0.
SR001, SKIP
Comments
• Channel(s) set to SKIP are not measured.
Setting voltage, thermocouple, RTD or contact input
Setting
SRp1, p2, p3<terminator>
p1
Channel number (001 to 030)
p2Type of input
VOLT DC voltage
TC
Thermocouple
RTD
Resistance temperature detector
DI
Contact
p3
Measurement range
p4
Span left value
p5
Span right value
Example
Set channel 01 of slot 1 to the thermocouple type R,and
set span left value to 0
°
C, span right value to 1760.0
°
C.
SR101, TC, R, 0, 1760.0
Comments
• For the p3 measurement range, see the range
parameter table on Pages 4-7 and 4-8.
• Set the span in p4 and p5 within the setting range
shown on pages 4-7 and 4-8.
•
For p4 and p5 setting value, enter them within 6
digits excluding the decimal point. The decimal point is
fixed. Refer to setting range on pages 4-7 and 4-8.
Setting DELTA(difference between channels)/
RRJC(remote RJC)
Setting
SRp1, p2, p3<terminator>
p1
Channel number (001 to 030)
p2DELTA/RRJC
p3
Reference channel (01 to 29)
p4
Span left value
p5
Span right value
Example
Set channel 10 of slot 2 to DELTA that is difference
value from channel 01 of the same unit,and set span
left value to -100.0, span right value to 100.0.
SR210, DELTA, 01, -100.0, 100.0
Comments
• As the reference channel, set a channel of the same
unit as the channel to be set to DELTA or RRJC. In
addition, the reference channel number must be
smaller than the source channel number.
• For RRJC, input for the reference channel must be
of a thermocouple type.
• If the input mode (type of input, scaling...) or the
measurement range of the reference channel is
changed, DELTA or RRJC setting is cleared. The
input mode changes to the mode before the reference
channel was changed.
• If any change has been made to the number, input
mode (type of input or scaling), or measuring range
of a given reference channel, the alarm setting of
the channel for which the item “DELTA/RRJC” is
being set changes to “OFF.”
• If any change has been made to the number or input
mode (type of input or scaling) of a given reference
channel, to the measuring range, or to the span of
the set channel, the partial-expansion setting of the
channel for which the item “DELTA/RRJC” is being
set changes to “OFF.”
• Set the span in p4 and p5 within the setting range
for the reference channel shown on pages 4-7 and
4-8.
• For p4 and p5 setting value, enter them within 6
digits excluding the decimal point. The decimal point
is fixed. Refer to setting range on pages 4-7 and 4-8.
• RRJC setup is valid only for recorders with the
optional computing function.
5.1 Setting the Input
Summary of Contents for DR130 SERIES
Page 39: ...4 9 Command Format 4 4 5 Input Range Parameter This page intentionally left blank ...
Page 40: ...4 10 4 5 Input Range Parameter This page intentionally left blank ...
Page 56: ...5 15 Commands 5 This page intentionally left blank 5 5 Other Settings ...
Page 71: ...6 9 Output Format 6 This page intentionally left blank ...