5-7
Commands
5
SZ
Sets the recording zone.
Mode
Operation mode
Setting
SZp1, p2, p3<terminator>
p1
Channel number (001 to 030)
Computation channel number (A01 to A30)
p2Lower limit of zone (0 to 2
45 mm, or 0 to 145
mm for the DR130)
p3
Upper limit of zone (5 to 250 mm, or 5 to 150
mm for the DR130)
Example
Set the recording zone of channel 6 in slot 3 to
10 to 50 mm.
SZ036, 10, 50
Comments
• Set p2 smaller than p3.
• This is effective for analog recording.
• The optional computation channel number for a
stand-alone model is A01 to A30.
SP
Sets partially expanded recording.
Mode
Operation mode
Setting
SPp1, p2, p3, p4<terminator>
p1
Channel number (001 to 030)
Computation channel number (A01 to A30)
p2ON/OFF of partially expanded recording
ON
Performs partially expanded recording.
OFF
Does not perform partially expanded
recording.
p3
Boundary recording position (1 to 99%)
p4
Boundary value
Example
Partially compress and record channel 6 in slot 3 at the
boundary recording position of 25% for the boundary
value 0 V.
SP036, 25, 0
Comments
• A waveform can be partially expanded by setting
the set input level (boundary value) as a percentage
of the recording chart (boundary recording position).
• This is effective for analog recording.
• If contiguous channels are set to p1, the decimal point
in boundary values when the measurement ranges
for set channels differ take the positions determined
for each measurement range (see Pages 4-7 and 4-
8). If the result exceeds the measurable ranges, an
error occurs. For example, if the channels of
measurement ranges 20 mV, 2 V, and type T
thermocouple are set to p1 and the boundary value
is set to 10000, the following applies:
• The boundary value for the channel of the 20 mV
measurement range: 10.000 mV
• The boundary value for the channel of the 2 V
measurement range: 1.0000 V
• The boundary value for the channel of the type T
thermocouple: 1000.0
°
C
Since the measurement range of the type T
thermocouple is -200.0 to 400.0
°
C, an error occurs.
For decimal point positions, see Pages 4-7 and 4-
8.
• The optional computation channel number for a
stand-alone model is A01 to A30.
• Set p4 to a value within the recording span, or within
the left/right scale if linear scaling is applied. It is
not possible, however, to set the recording span nor
the left/right scale.
• In the case of power monitor module, if the settings
of the wiring method and the input range are
changed, the partially expanded recording setting is
set to OFF.
SG
Sets or copies a message.
Message setting
Mode
Operation mode
Setting
SGp1, p2<terminator>
p1
Message number (01 to 20)
p2Message character string (Up to 16 characters)
Example
Set the message “test” to message No. 5.
SG05, test
Comments
• For details of the character strings that can be used
in messages, see the characters in the table of ASCII
codes on Page 4-11.
Message copy
Mode
Operation mode
Setting
SGp1, p2, p3<terminator>
p1
COPY
p2Message number at a message source (01 to
20)
p3
Message number at a message destination (01
to 20)
Example
Copy message No. 5 to message No. 12.
SGCOPY, 05, 12
ST
Sets a tag.
Mode
Operation mode
Setting
STp1, p2<terminator>
p1
Channel number (001 to 030)
Computation channel number (A01 to A30)
p2Tag character string (up to 16 characters)
Example
Set “TEST” as the tag for channel 6 in slot 3.
ST036, TEST
Comments
• For details of the character strings that can be used
as tags, see the characters in the table of ASCII codes
on Page 4-10.
• The optional computation channel number for a
stand-alone model is A01 to A30.
SH
Sets a header.
Mode
Operation mode
Setting
SHp1, p2<terminator>
p1
Setting line number (1 to 5)
p2Header character string
Example
Set the line of number 2 to “TEST RECORDE.”
SH2, TEST RECORDE
Comments
• Set the character string to be printed in the first place
of a record.
• For details of the character strings that can be used
as headers, see the characters in the table of ASCII
codes on Page 4-9.
• Parameter p2 can have no more than 60 characters
for the DR130 and no more than 80 characters for
the DR230/240.
SJ
Sets a title.
Mode
Operation mode
Setting
SJp1<terminator>
p1
Title character string (up to 32 characters)
Example
Set the title to “TEST RECORDE.”
SJTEST RECORDE
Comments
• Sets a character string which is recorded in a fixed
interval in analog recording. The interval to be
recorded is set using the XR command (see 5-9).
• For details of character strings that can be used as
titles, see the characters in the table of ASCII codes
on Page 4-9.
SF
Sets the format for list printing.
Mode
Operation mode
Setting
SFp1, p2, p3, p4<terminator>
p1
ON/OFF of range list printing (ON, OFF)
p2ON/OFF of alarm list printing (ON, OFF)
p3
ON/OFF of parameter list related to printing
(ON, OFF)
p4
ON/OFF of other parameters (ON, OFF)
5.4 Setting the Recording
Summary of Contents for DR130 SERIES
Page 39: ...4 9 Command Format 4 4 5 Input Range Parameter This page intentionally left blank ...
Page 40: ...4 10 4 5 Input Range Parameter This page intentionally left blank ...
Page 56: ...5 15 Commands 5 This page intentionally left blank 5 5 Other Settings ...
Page 71: ...6 9 Output Format 6 This page intentionally left blank ...