The previous figure can be used to understand how SSH is used in the DBK17 SSH Card. The process is
as follows:
•
Input signals pass through an instrumentation amplifier and into a sample-and-hold stage.
•
When the sample enable line goes high, each channel’s sample-and-hold stage will “freeze”
the current analog value. The values for all channels are separately “latched” within 50 ns of
each other.
•
The signals are held in a stable condition, while the multiplexer switches through all
channels.
•
The multiplexer sends the signals [one-by-one] to be digitized by the analog-digital converter
(ADC) in the primary data acquisition device.
•
The resulting data is a snapshot of conditions at an instant, but the multiplexing and analog-
to-digital conversion are spread over a longer time interval.
•
The simultaneous sample and hold circuit allows you to gather up to 256 simultaneous
samples using sixty-four DBK17s.
Two-Point Calibration of a Temperature Measurement System
Note
: In the following text a temperature measurement system consisting of a DaqBook, a DBK19
thermocouple card, and DaqView is referred to. However, it is important to note that the two-point
calibration method can be applied to measurement systems that consist of other acquisition devices
and software.
Overview
Two-point calibration makes use of correction constants
to eliminate offset and gain errors in a temperature
measurement system.
The graph shows an example for a temperature
measuring system, with temperature expressed in
degrees Celsius.
The sensor output is on the horizontal axis and system
response is on the vertical.
The graph illustrates the following bulleted points:
0
10
20
30
40
50
60
70
80
90
100
10
20
30
40
50
60
70
80
90
100
Temperature Sensor Output ( C)
°
Me
as
u
rem
en
t S
y
s
tem
Ou
tp
ut (
C
)
°
•
The system has a linear response. Most thermocouple measurement devices have a software
driver, which corrects for the nonlinear output voltage of each thermocouple type. Normally,
there is no provision for linearity adjustment by the user.
•
The system has an offset at 0°C, which can be eliminated by introducing an offset correction. This
is the first point of a two-point correction.
•
The system has a gain error at 100°C, which can be eliminated by introducing a gain (slope)
correction. This is the second point of a two-point correction.
Suppose you have a temperature measurement system consisting of a DaqBook, a DBK19 thermocouple
card, and DaqView data acquisition software. The DBK19 was shipped with a diskette of calibration
constants. These calibration constants are specific to the DBK19.
1-12
Signal Management
886995
DBK Option Cards and Modules
Summary of Contents for OMB-DBK-34A
Page 6: ...ii...
Page 10: ...viii 917594 DBK Option Cards Modules User s Manual This page is intentionally blank...
Page 32: ...pg 22 DBK Basics 967794 Daq Systems...
Page 60: ...2 10 System Connections Pinouts 877095 DBK Option Cards and Modules...
Page 84: ...5 8 Troubleshooting Tips 967094 DBK Option Cards and Modules...
Page 94: ...DBK200 Series Boards DBK200 DBK201 DD 10 949794 Dimensional Drawings...
Page 96: ...DBK205 DBK205 DD 12 949794 Dimensional Drawings...
Page 97: ...DBK206 Dimensional Drawings 949794 DD 13...
Page 99: ...DBK208 Dimensional Drawings 949794 DD 15...
Page 100: ...DBK209 DD 16 949794 Dimensional Drawings...
Page 101: ...Dimensions for Miscellaneous Components DIN 1 Dimensional Drawings 949794 DD 17...
Page 102: ...DIN 2 DD 18 949794 Dimensional Drawings...
Page 105: ...DBK Cards Modules Part 1 of 2...
Page 106: ...DBK Cards Modules...
Page 108: ...DBK Cards Modules...
Page 148: ...DBK7 pg 14 879895 DBK Option Cards and Modules...
Page 168: ...DBK15 pg 6 889094 DBK Option Cards and Modules...
Page 182: ...DBK16 pg 14 879895 DBK Option Cards and Modules...
Page 200: ...DBK16 pg 32 879895 DBK Option Cards and Modules...
Page 206: ...DBK17 pg 6 879895 DBK Option Cards and Modules...
Page 218: ...DBK20 and DBK21 pg 6 879795 DBK Option Cards and Modules...
Page 232: ...DBK24 pg 8 879795 DBK Option Cards and Modules...