Shunt Calibration
Shunt calibration is virtually identical to the 2-Point method just discussed, except that the second point is
simulated. The simulated load is achieved by shunting one leg of the bridge with a shunt resistor. Shunt
calibration is the preferred calibration method when applying a real load (of a value approximating the
maximum expected load) is not practical. To adjust the channel gain, the shunt must be applied to the
bridge.
Calculate and install the necessary shunt resistor before continuing.
•
DBK43A
has direct support for shunt calibration, accommodating the
resistor in its enclosure and allowing the software to apply it when requested.
•
DBK16
does not have direct support, so the shunt resistor must be applied
externally and switched in manually.
Adjust the Offset
Adjust the offset as follows.
1.
In the Param1 column, select all of the DBK43A channels that are to be adjusted.
2.
Select
Mode = Bridge
from the drop down list above the grid. This selection commands the
calibration multiplexer to route the transducer voltage through the analog path.
3.
Turn off all the channels in the system except for those DBK43A channels that are to be
adjusted.
4.
Click the Download button to send the current configuration to the LogBook.
5.
Select Indictors \ Enable Input Reading Column from the menu bar to display the offset
values for each channel.
6.
For the associated channel, apply the first calibrated load to each gage (typically no-load) and
set the voltage to 0.0V for each transducer by adjusting the trimpot labeled OFFSET.
If the first point is an actual calibrated load, you must move the load to each gage, one at a
time, to adjust its associated offset.
Adjust the Input and Scale Amplifier Gain
For the best results, a gain should be selected so that the bridge’s maximum output equals 90% of the
A/D’s input range.
1. Use the following equation to calculate the desired shunt voltage (V
D
).
V
D
= (L
s
/L
M
) * V
I
* 90%
Where:
V
D
= Desired voltage from the after amplification when the shunt is
applied
L
s
= Simulated load produced by shunt
L
M
= Maximum load expected during usage
V
I
= Input voltage range
Example:
The simulated load produced by the shunt 100lbs. The maximum expected load
during usage is 150lbs. The programmable input range of the data acquisition system is set for
+5V. The desired output voltage of the strain gage signal conditioning electronics is:
V
D
= (100/150) * 5 * 0.90 = 3V
In this example, we would adjust the GAIN and SCALE trimpots until a value of 3V is
measured.
If 150lbs is applied to the gage, a voltage of 4.5V will be measured.
DBK16, pg. 28
879895
DBK Option Cards and Modules
Summary of Contents for OMB-DBK-34A
Page 6: ...ii...
Page 10: ...viii 917594 DBK Option Cards Modules User s Manual This page is intentionally blank...
Page 32: ...pg 22 DBK Basics 967794 Daq Systems...
Page 60: ...2 10 System Connections Pinouts 877095 DBK Option Cards and Modules...
Page 84: ...5 8 Troubleshooting Tips 967094 DBK Option Cards and Modules...
Page 94: ...DBK200 Series Boards DBK200 DBK201 DD 10 949794 Dimensional Drawings...
Page 96: ...DBK205 DBK205 DD 12 949794 Dimensional Drawings...
Page 97: ...DBK206 Dimensional Drawings 949794 DD 13...
Page 99: ...DBK208 Dimensional Drawings 949794 DD 15...
Page 100: ...DBK209 DD 16 949794 Dimensional Drawings...
Page 101: ...Dimensions for Miscellaneous Components DIN 1 Dimensional Drawings 949794 DD 17...
Page 102: ...DIN 2 DD 18 949794 Dimensional Drawings...
Page 105: ...DBK Cards Modules Part 1 of 2...
Page 106: ...DBK Cards Modules...
Page 108: ...DBK Cards Modules...
Page 148: ...DBK7 pg 14 879895 DBK Option Cards and Modules...
Page 168: ...DBK15 pg 6 889094 DBK Option Cards and Modules...
Page 182: ...DBK16 pg 14 879895 DBK Option Cards and Modules...
Page 200: ...DBK16 pg 32 879895 DBK Option Cards and Modules...
Page 206: ...DBK17 pg 6 879895 DBK Option Cards and Modules...
Page 218: ...DBK20 and DBK21 pg 6 879795 DBK Option Cards and Modules...
Page 232: ...DBK24 pg 8 879795 DBK Option Cards and Modules...