21
IX73
c
d
Movement range limit screw
As a factory default, IX3-SVR, is equipped with the movement range
limit screw which limits the stage movable range in the vertical or
horizontal direction.
To enlarge the movable range to observe microplates, etc., remove
the movement range limit screw.
Remove the movement range limit screw
e
in the vertical direction
with the Allen screwdriver provided with the microscope.
Rotate the longitudinal handle
f
to move the stage inward.
Remove the movement range limit screw
g
in the horizontal direc-
tion with the Allen screwdriver provided with the microscope.
1
2
3
e
g
f
b
a
•
Do not attempt to rotate the stage handle forcedly exceeding
the stage movable range. The stage may be damaged.
•
As the objective may interfere with the stage depending on the
focus position, be sure to operate carefully.
CAUTION
Mechanical stage with right handle (IX3-SVR)
To move the specimen to a desired position, rotate the X-axis knob
a
and Y-axis knob
b
.
} If the stage is used for a long time period, the stage movement
range may be narrower rarely. In this case, move the stage several
times within the full movement range toward the front/back or right/
left direction while holding the top surface of the stage with both
hands.
Movement control knob
Attaching the movement control knob provided with IX3-SVR makes
it difficult to move the stage in the blocked direction. Even though
you may touch the stage accidentally during observation, the obser-
vation position can be secured.
If the movement control knob is attached to the hole of
c
, the
Y-axis movement will be blocked. If it is attached to the hole of
d
(backside of the stage), the X-axis movement will be blocked.
Do not tighten the movement control knob too firmly. The stage
may be damaged.
} If you tighten the movement control knob while focusing on the
specimen, it will be defocused.
CAUTION
2
Moving the Specimen