52
a
b
c
d
d
Aligning the two guide holes on the illumination column with the
two protruding positioning pins on the microscope, gently fit the
column onto the microscope from the above.
While holding the illumination column with one hand, insert the four
provided Allen screws into the screw holes. Then tighten the screws
with the provided Allen wrench.
Do not tighten the screws with the illumination column inclined.
Doing so may damage the screw sections.
1
2
CAUTION
} The designated bulb model is the 12V100WHAL-L (PHILIPS 7724)
halogen bulb.
Fully loosen the clamping screw at the top of the halogen lamp
housing using the Allen screwdriver provided with the microscope.
Lift up the halogen lamp housing to remove it.
Tilt the bulb socket by 90° in the direction of the arrow.
While pushing down the bulb clamping lever
a
, hold the halogen
bulb
b
with gloves or a piece of gauze, insert the bulb pins
c
straight and fully into the pin position
d
on the lamp socket.
Then return the lamp clamping lever gently back to the original
position to clamp the bulb.
1
2
3
4
To prevent reduced bulb life or cracking, do not touch the bulb
with bare hands. If fingerprints are accidentally left on the bulb,
wipe the bulb with a dry soft cloth.
Fit the halogen lamp housing from up and tighten the clamping
screw by applying downward pressure.
Caution for Bulb Replacement During or Right After Use
The bulb, lamp housing and areas near these will be extremely
hot during and right after use.
Set the main switch to “
\
” (OFF), disconnect the power cord
from the wall outlet, then allow the old bulb and lamp housing
to cool before replacing the bulb with a new of the designated
type.
CAUTION
5
2
Mounting the Illumination Column
3
Attaching the Halogen Bulb
1
2
1
5
2
3