Loosen the cover fixing screw
a
(2 positions) by using the Phillips
screwdriver, and remove the cover
b
.
1
Insert the revolving nosepiece slowly into the positioning pins
c
(2
positions) in the upper area of the microscope.
As the mounting holes of the positioning pins are small, insert
the revolving nosepiece into the positioning pins carefully from
the above paying attention not to tilt the revolving nosepiece.
2
CAUTION
Tighten the revolving nosepiece fixing screws
d
(5 positions) by using
the Allen wrench provided with the microscope.
3
Place the cover
b
over the revolving nosepiece, and tighten the
cover fixing screws
a
(2 positions) by using the Phillips screwdriver.
4
a
b
c
d
a
b
To ensure the performance, we recommend that you have your
Olympus representative to assemble the revolving nosepiece.