57
IX73
1
1
2
a
Remove the stage center plate and attach the objectives to the
revolving nosepiece through the hole on the stage left by the plate.
} Attach the objectives in such a manner that the magnification
increases from low to higher powers in a clockwise direction.
} Use the hand switch U-HSCBM to rotate the motorized 6 position
nosepiece IX3-D6REA.
Do not rotate the motorized revolving nosepiece manually. Oth-
erwise, it will cause damage.
In the inverted microscope, the front lens of the objectives
faces upward, and is more exposed to contamination than the
objectives of upright microscopes. Therefore, if there are empty
positions
a
in the nosepiece, attach the dust caps provided.
After all objective lenses are mounted, fit the stage center plate into
the stage.
1
CAUTION
CAUTION
2
If the condenser refocusing stopper is attached when the
condenser attaching arm is placed at the lower position, the
condenser may not be attached. Be sure to raise the condens-
er attaching arm before attaching the condenser refocusing
stopper.
} Mount the condenser refocusing stopper in the direction shown in
the picture.
Insert the condenser refocusing stopper between the column and
the condenser attaching arm.
Secure the condenser refocusing stopper with the fixing screw by
using the Allen screwdriver provided with the microscope.
Be careful not to drop the fixing screws.
CAUTION
1
2
CAUTION
7
Attaching the Objectives
8
Attaching the Condenser Refocusing Stopper