10-015490-01EN, Rev. 2, June 2020
Chapter 3
48
3.6
Unloading a Sample
To remove the sample carrier from the sample chamber
1.
Lift the lever to unlock the sample carrier.
2.
Lift the sample carrier out of the sample chamber.
To unload a sample from the sample cell
Remove the majority of the material with the aid of the external shaker assembly.
OR
1.
Using a 1.5 mm hex driver, carefully remove the holding screws, and then
disassemble the cell windows and spacer (see Figure 3-17 on page 48).
2.
Clean or replace the cell parts as necessary.
Figure 3-17 Cell disassembly