10-015490-01EN, Rev. 2, June 2020
Set Up and Operation
43
To load a sample into a
sample
cell
1.
Connect the external shaker plug to the External Shaker connector on the rear
panel (see Figure 3-13 on page 43).
Figure 3-13 External shaker plug connected
2.
Insert the sample carrier into the external shaker and gradually load a small
amount of material (enough to fill the gap created by the spacer) into the side A
cell assembly (see Figure 3-14 on page 44).
The material should be approximately 50 mg of coarsely ground (100–150 µm)
powder.
External shaker plug