10-015490-01EN, Rev. 2, June 2020
Safety Information
31
Figure 2-1 BTX III radiation safety features
Table 5 BTX III radiation safety features
Label
Safety feature
1
Sample chamber interlock
The X-ray beam only operates when the sample carrier is properly
inserted into the system. If the carrier is removed while the system is in
operation, the high-voltage power supply interrupts and X-ray
generation ceases.
High Voltage/X-ray On LEDs
Interlock open LED
4
1
2
3