DMTA-10022-01EN, Rev. C, January 2015
Chapter 6
88
6.6
High-Penetration Software Option
The 45MG high-penetration software option coupled with low-frequency single
element transducers (down to 0.5 MHz) allows you to make thickness, material sound
velocity, and time-of-flight measurements on materials such as composites, fiberglass,
plastic, rubber, and cast metals that are difficult or impossible to measure using
standard ultrasonic instruments. The M2008 [U8415001] transducer is a special low-
frequency transducer used to measure the thickness of thick fiber-reinforced polymer
(FRP) and composite materials.
With a M2008 transducer only, press
[2nd F]
,
[CAL ZERO] (Do ZERO)
at any time to
auto-adjust the zero offset and compensate for temperature changes in the delay line.
To use the high-penetration software option with the M2008 transducer
1.
Ensure that the high-penetration software option is activated (see “Activating
Software Options” on page 68 for details).
2.
Connect the M2008 transducer to the T/R 1 connector at the top of the 45MG.
3.
Press
[2nd F]
,
[FREEZE] (XDCR RECALL)
.
4.
In the menu, select
DEFAULT HP SINGLE ELEMENT
.
5.
In the
DEFAULT HP SINGLE ELEMENT
screen, highlight the default transducer
setup for the M2008 (
DEFP1-0.5-M2008
), or any custom setup that uses the M2008
transducer.
6.
Press
[MEAS]
to return to the measurement screen with the recalled transducer
setup.
7.
Wipe the couplant off the transducer tip.
8.
Press
[2nd F]
,
[CAL ZERO] (Do ZERO)
.
9.
Perform the material sound velocity and zero calibration (see “Calibrating the
Instrument” on page 55).
6.7
Datalogger Option
This section describes how to use the 45MG internal datalogger to organize your data.
NOTE
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...