DMTA-10022-01EN, Rev. C, January 2015
Chapter 5
52
Figure 5-1 Plugging in the transducer
2.
Press
to start the instrument.
The measurement screen appears. With a standard D79X dual element
transducer, the “
Do--
” message appears in the measurement screen (see
Figure 5-2 Initial screen with a standard D79X dual element transducer
3.
For dual element transducers, perform the transducer zero compensation:
a)
Wipe the couplant off the tip of the transducer.
b)
Press
[2nd F]
,
[CAL ZERO] (Do ZERO)
.
T/R 1 connector for a
single element transducer
Standard dual element
transducer connector