DMTA-10022-01EN, Rev. C, January 2015
Maintaining and Troubleshooting the 45MG
205
7.
When
DUAL XDCR TEST
is selected, in the
DUAL XDCR TEST
screen (see
a)
Press
[MEAS]
to start the test that measures the time of flight through each
side of a dual element transducer.
The measured parameter values appear.
b)
Interpret the
Tx
and the
Rx
values as follows:
•
Similar values indicate a normal dual element transducer.
•
A difference between the values indicates that the delay line wear is not
the same for each element.
•
A missing value indicates that a cable is broken, or that the element is not
functioning.
c)
Take note of the calculated
ZERO VALUE
.
d)
Press
[ENTER]
to terminate the dual element transducer test.
Figure 12-3 The DUAL XDCR TEST screen
12.6 Performing the Software Diagnostic Test
The software diagnostic (
SW DIAG
) function generates an error log documenting all
errors that occurred during instrument operation. Olympus uses this information to
troubleshoot the 45MG operating software.
To access the software diagnostic
1.
In the measurement screen, press
[SETUP]
, and then select
SP MENU
.
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...