DMTA-10022-01EN, Rev. C, January 2015
Maintaining and Troubleshooting the 45MG
203
Figure 12-1 Activating a reset function
12.5 Performing Hardware Diagnostic Tests
The 45MG includes a function that can be used to perform several diagnostic self
tests. The tests can help to localize a suspected hardware problem, or to verify correct
hardware operation. Some tests are designed for Olympus internal testing procedures
during the manufacturing process.
The available tests are as follows:
•
Keypad test
•
Video test
•
Internal microSD card test (Pass or Fail)
•
External microSD card test (Pass or Fail)
•
Dual element transducer test
•
ESS test (Electronic Stress Screening) [for Olympus internal use only]
•
B-scan test (for Olympus internal use only)
•
Battery test (for Olympus internal use only)
•
One-wire test (for Olympus internal use only)
To perform a diagnostic test
1.
In the measurement screen, press
[SETUP]
, and then press
SP MENU
.
2.
In the
SP MENU
(see Figure 4-2 on page 45), select
TESTS
.
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...