DMTA-10022-01EN, Rev. C, January 2015
Using Special Functions
129
Figure 7-2 The DIFF screen
When
DIFF MODE
is set to
REDUCTION RT
only:
d)
In the
FORMER THICKNESS
text box, enter the original thickness value, as
measured before bending the metal.
e)
In the
LARGE FONT text
box, select which measurement is to appear at the
bottom of the measurement screen with the large type font (
THICKNESS
or
REDUCTION RT
).
3.
Press
[MEAS]
to return to the measurement screen with the displayed differential
value.
7.2
Using the Minimum, Maximum, or Min/Max Thickness Mode
The minimum, maximum, or min/max thickness modes can be activated to display
retained minimum and/or maximum thickness values. The
MIN
and/or
MAX
values
appear on the left side of the main thickness reading (see Figure 7-3 on page 130). The
minimum or maximum thickness value replaces the main thickness display when the
transducer is uncoupled, or when a loss of signal (LOS) occurs. The replacement value
appears in an outlined font.
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...