DMTA-10022-01EN, Rev. C, January 2015
Chapter 9
152
To use the extended blank
1.
Press
[GAIN/WAVE ADJ]
(only available when the waveform is active).
The waveform adjustment parameter and its value appear on the measurement
screen (see Figure 9-2 on page 152).
2.
If needed, use the
[
]
and
[
]
keys to select
EXT BLANK
.
The extended blank becomes active, but initially, its value is zero. The gage
remains in the measurement screen.
3.
Use the
[
]
or
[
]
keys respectively to increase or decrease the blank value until
early unwanted echoes are excluded from detection.
The extended blank horizontal bar just above the waveform display indicates the
extended blank length (see Figure 9-2 on page 152).
Figure 9-2 Adjusting the extended blank length
4.
To turn off the extended blank, use the
[
]
key to bring the extended blank value
back down to zero.
Waveform adjustment
parameter and value
Extended blank bar
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...