DMTA-10022-01EN, Rev. C, January 2015
Chapter 10
168
c)
Select
SAVE
.
8.
Press
[MEAS]
to return to the measurement screen.
The saved setup becomes the active setup.
10.2 Quickly Adjusting Waveform Parameters For Single Element
Transducers
With single element transducers, you can quickly adjust individual waveform
parameters using the
[GAIN/WAVE ADJ]
key.
To quickly adjust individual waveform parameters
1.
Ensure that a single element transducer is connected to the 45MG.
2.
In the measurement screen, press
[GAIN/WAVE ADJ]
.
The waveform adjustment parameter appears above the thickness value on the
measurement screen (see Figure 10-2 on page 168).
Figure 10-2 Adjusting the waveform parameters
3.
Use the
[
]
or
[
]
key to select the parameter that you want to adjust. The choices
are as follows:
•
DET MODE
(see “Detection Modes” on page 169 for details)
•
M3 BLANK
in mode 3 only (see “Mode 3 Echo Blank” on page 182 for
details)
Waveform parameter
value
Waveform parameter
Summary of Contents for 45MG
Page 8: ...DMTA 10022 01EN Rev C January 2015 Table of Contents viii List of Tables 227 Index 229...
Page 26: ...DMTA 10022 01EN Rev C January 2015 Introduction 16...
Page 36: ...DMTA 10022 01EN Rev C January 2015 Chapter 1 26...
Page 42: ...DMTA 10022 01EN Rev C January 2015 Chapter 2 32...
Page 136: ...DMTA 10022 01EN Rev C January 2015 Chapter 6 126...
Page 150: ...DMTA 10022 01EN Rev C January 2015 Chapter 7 140...
Page 157: ...DMTA 10022 01EN Rev C January 2015 Configuring the Instrument 147 Figure 8 3 The COMM screen...
Page 158: ...DMTA 10022 01EN Rev C January 2015 Chapter 8 148...
Page 196: ...DMTA 10022 01EN Rev C January 2015 Chapter 10 186...
Page 208: ...DMTA 10022 01EN Rev C January 2015 Chapter 11 198...
Page 226: ...DMTA 10022 01EN Rev C January 2015 Appendix A 216...
Page 232: ...DMTA 10022 01EN Rev C January 2015 Appendix C 222...