Figure 7-1b: Technicians Menu Tree- Wireless Models
7.2 Set Balance / Set Sensor Voltage (Technicians only!)
Set Balance
and
Set Sensor Voltage
are used when
Input Type
is for
Bridge
sensors. They are
factory configured
and only require field adjustment if the catalytic
bead sensor is mounted remote from the OLCT 200 or if a new sensor is installed.
Other input type entries draw a line through these menus and they are inactive. OLCT
200 bridge sensors may be from 2 to 6 volts excitation voltage
at the sensor
. This
means if the sensor is mounted a long distance away the voltage at the OLCT 200 may
have to be higher to compensate for losses in field wiring. Be careful not to exceed
correct sensor volts at the
sensor’s A and R terminals.
Set Balance
allows balancing of the catalytic bead sensor and must only be performed
with ZERO gas on the sensor (Figure 7-2). Balance is similar to a very coarse ZERO
calibration and does not need to be precise since subsequent calibrations will correct
for small errors. ZERO gas applied to the sensor should provide a Reading of
–3 to +3
on the SENSOR BALANCE menu.
Summary of Contents for OLCT 200
Page 4: ...OLCT 200 User Manual 4...
Page 14: ...OLCT 200 User Manual 14 Figure 2 3a Outline Drawing OLCT 200 Wired Models...
Page 19: ...OLCT 200 User Manual 19 2 6 2 Installation Drawing 11 0100...
Page 32: ...OLCT 200 User Manual 32...
Page 36: ...OLCT 200 User Manual 36...
Page 38: ...OLCT 200 User Manual 38 Figure 4 1 Calibration Gas Input...
Page 61: ...OLCT 200 User Manual 61 Figure 6 5...
Page 62: ...OLCT 200 User Manual 62...
Page 79: ...OLCT 200 User Manual 79...