OLCT 200 User Manual |
54
Chan 2 Alarm 1 Real
41031
4
NA
Chan 2 Alarm 1 Devisor 41032
4
NA
Chan 2 Alarm 2 Real
41033
4
NA
Chan 2 Alarm 2 Devisor 41034
4
NA
Chan 2 Alarm 3 Real
41035
4
NA
Chan 2 Alarm 3 Devisor 41036
4
NA
Chan 2 Manual Gain Real
41037
4
NA
Chan 2 Manual Gain Devisor 41038
4
NA
Chan 2 Manual Offset Real
41039
4
NA
Chan 2 Manual Offset Devisor 41040
4
NA
Binary Cal Data:
Chan 1 A2D MIN
41041
4
NA
Chan 1 A2D MAX
41042
4
NA
Chan 1 D2A MIN
41043
4
NA
Chan 1 D2A MAX
41044
4
NA
Chan 2 A2D MIN
41045
4
NA
Chan 2 A2D MAX
41046
4
NA
Chan 2 D2A MIN
41047
4
NA
Chan 2 D2A MAX
41048
4
NA
Min and Max calibration points for the A/D and D/A converters.
5.11 SYSTEM SECURITY
The
SYSTEM
SECURITY
menu offers two levels of protection. A
LOW
level allows
CAL MODE sensor calibrations but requires the 4-digit
Pass Code
prior to altering
menus.
HIGH
level locks the entire menu database and the CAL Mode until the correct
Pass
Code
is entered.
LOW
and
HIGH
security levels always allow viewing of
configuration menus but they may not be changed.
Contact
Name
is a 12 character
ASCII field available for displaying a phone # or name of personal who know the
Pass
Code
. Lost
Pass Codes
may be recovered by entering the locked security menu and
holding the UP key for 5 seconds. The 4-digit code appears near the bottom of the
screen.
Figure 5-13: System Security Menu
S Y S T E M S E C U R I T Y
C o n t a c t N a me
S e c u r e L e v e l L OW
P a s s C o d e
U n l o c k e d
Summary of Contents for OLCT 200
Page 4: ...OLCT 200 User Manual 4...
Page 14: ...OLCT 200 User Manual 14 Figure 2 3a Outline Drawing OLCT 200 Wired Models...
Page 19: ...OLCT 200 User Manual 19 2 6 2 Installation Drawing 11 0100...
Page 32: ...OLCT 200 User Manual 32...
Page 36: ...OLCT 200 User Manual 36...
Page 38: ...OLCT 200 User Manual 38 Figure 4 1 Calibration Gas Input...
Page 61: ...OLCT 200 User Manual 61 Figure 6 5...
Page 62: ...OLCT 200 User Manual 62...
Page 79: ...OLCT 200 User Manual 79...