65
UNO-T-H-CO2
–
Technical Specifications
Temperature
Control Range: 3°C above ambient temperature (minimum temperature set point 25°C) to 60°C
Step size: 0.1°C
Accuracy:
•
± 0.1°C in Sample Control mode
•
± 0.3°C in Chamber Control mode, if the Room Temperature is stable within ± 1°C.
Pressure Gauge
–
input gas
CO2 gas supply
The CO2 tank must be equipped with a pressure regulator to decrease the pressure from the
pressure tank to 2 barg (29 psig).
Pressure Gauge
–
input gas
connector
Push to fit connector for 6 mm O.D. rigid tube
2GF-MIXER
CO2 Control Range in air: 0 -15 %
Accuracy: ±1%
CO2 flow rate Range: 0.013
–
0.13 l/min
CO2 inlet connector: Push to fit connector for 6 mm O.D. rigid tube
Air flow rate Range: 0.1
–
1.0 l/min
Air inlet connector: Push to fit connector for 6 mm O.D. rigid tube
Gas outlet connector: Push to fit connector for 6 mm O.D. rigid tube
OKO-AP Outlet pressure
1.3 bar (19 psi)
OKO-AP Outlet flow rate
•
Low Flow: 0
–
0.5 l/min
•
High Flow: 0.5
–
1.0 l/min
Humidification Power
85% - 90% Relative Humidity
if the Room Temperature is stable within ±1°C at 23°C, the GF-MIXER-HM (Humidity Module) is
installed away from heating and air conditioning vents and windows /doors that are frequently
open.
Power Supply
UNO-CONTROLLER
–
Supply Voltage 24 V-DC
GF-MIXER-HM (Humidity Module)
–
Supply voltage
24 V DC
OKO-AP - Supply Voltage
12 V-DC
Power Consumption
120 W max
Dimension
UNO-CONTROLLER (LxWxH)
129mmx125mmx104mm
UNO-CONTROLLER Weight
0.400 Kg
OKO-AP (LxWxH)
96mmx160mmx57mm
OKO-AP Weight
0.350 Kg
Humidifier Bottle
diameter 72 mm, height 190 mm
2GF-MIXER (LxWxH)
126mmx200mmx178mm
2GF-MIXER Weight
3.0 Kg
Table 7. UNO-T-H-CO2 Technical Specifications