40
a
b
Figure 54. Settings
–
Calibrations - Chamber
–
Log (a
–
b).
UNO-CONTROLLER stores the parameters obtained by previous calibrations for each combination of
selected chamber with its selected sample holder. You can recover calibration parameters only for
the selected pair of chamber and its sample holder.
2.
By pressing
Factory
button (see pointer 2 in Figure 53 b) UNO-CONTROLLER will recover and use the
factory calibration parameters.
Note
►
Factory calibration was performed at a Room Temperature of 23°C for a temperature set point of
37°C. Use factory calibration parameters if your Room Temperature is 23°C ±1°C.
3.
By pressing
Calibrate
(see pointer 3 in Figure 53 b) UNO-CONTROLLER will start a routine to calibrate its
parameters using the T Sensor.
Calibrate UNO-CONTROLLER if you Room Temperature differs by more than ±1°C from 23°C.
To start the chamber calibration, follow the instructions below:
•
Make sure that T Sensor is connected to UNO-CONTROLLER (see paragraph 7.4).
•
Secure with commercial adhesive tape the T Sensor to the bottom of the sample holder you use during
standard operation (Petri dish, Chambered Glass Slide, etc.) as shown in Figure 55. Do not cover the
tip of the T-Sensor with the adhesive tape.