8.2.3
Objective Heater
Press on
Objective
Heater
icon (see Figure 43 a) to enter the Objective Heater menu, see Figure 43 b.
Note
►
The Objective Heater is enabled by default. Disable the Objective Heater if it is not in use; otherwise,
UNO-CONTROLL will be in alarm.
a
b
1
2
3
Figure 43. Ho to enter the Obj, Heater menu (a -b).
At this point you can choose:
1.
By pressing on
Ramp Rate
button, the submenu in Figure 44 b will appear. Use the slide bar or the
+
and
–
icons to select the maximum heating rate (in °C/min) for the Objective Heater and press
Save
to confirm.
a
b
1
2
Figure 44. Settings
–
Obj. Heater
–
Ramp rate (a
–
b).
Tip
►
It is particularly indicated to use a low heating rate in case of Objective Heater acting on a very
sensitive objective. Check with the objective manufacturer.
Note
►
The ramp rate can be defined in the range 0.1-1.5°C/min. Above 1.5°C UNO-CONTROLLER attributes
the value MAX to the ramp rate. If MAX is selected, UNO-CONTROLLER will operate without constraint on the
maximum heating rate for the Objective Heater.
2.
Press
Cancel t
o discard your choice (see pointer 2 in Figure 43 b).
3.
Press
OK
to confirm (see pointer 3 in Figure 43 b), the menu in Figure 45 b will appear.