1
1
Preface
UNO-STAGE-TOP-INCUBATOR
ensures the highest temperature accuracy on sample temperature by
controlling simultaneously up to six integrated channels with a state-of-art multichannel controller.
UNO-STAGE-TOP-INCUBATOR is designed to control the environmental conditions inside the top stage
incubator, where specimens are located. UNO-STAGE-TOP-INCUBATOR features a system to control the
temperature from 3°C above ambient temperature (minimum temperature set point 25°C) to 60°C. The 4.3'' touch
screen controller interface allows working in both Chamber and Sample control mode. The temperature accuracy
is ± 0.1°C in Sample Control mode and ± 0.3°C in Chamber Control mode, if Room Temperature is stable within ±
1°C. UNO-STAGE-TOP-INCUBATOR includes an external fine gauge thermocouple for sample feedback operation.
UNO STAGE-TOP-INCUBATOR requires the selection of one stage top chamber. It is compatible with all
Okolab H301 incubating chamber (to be ordered separately).
UNO-STAGE-TOP-INCUBATOR is available in three different configurations:
•
UNO-T
including:
o
UNO-CONTROLLER
- the temperature controller.
•
UNO-T-H-PREMIXED
, stage top incubator's controller for application with premixed gas, including
o
UNO-CONTROLLER
- the temperature controller.
o
GF-REGULATOR
- flow rate regulator for the premixed gas. The regulator, equipped with a calibrated
output orifice, allows to set the desired output gas flow rate (in the range 0.1 - 0.4 l/min) by regulating
manually the pressure.
o
GF-MIXER-HM
- the Humidity Module. Dry premixed gas is humidified by flowing over warm water, heated
up by the Humidifier. The Humidifier is composed of a glass bottle, Humidifier Bottle, and its heater A
heated tube delivers the humid premixed gas to the stage top incubator preventing moisture
condensation.
•
UNO-T-H-CO2
, stage top incubator's controller for application with 100% CO2 and background air, including:
o
UNO-CONTROLLER
- the temperature controller.
o
GF-MIXER-HM
- the Humidity Module.
o
2GF-MIXER
- a two-channel gas flow regulator. The 2G-MIXER is a manual mixer that allows the mixing of
CO2 and air at the desired concentration by two floating ball flow meters. The CO2 concentration can be
regulated in the range 0-15% in air with an accuracy of ±1%.
o
OKO-AP
- air pump. It enables to use background air as a convenient alternative to the compressed air.
UNO-STAGE-TOP-INCUBATOR is also compatible with Okolab objective Heater, OBJ-COLLAR series,
recommended for immersion objectives. Collars connect directly to the controller and are calibrated with
automatic routine.
The controller features on-board memory for data logging and mini-USB port for data download.
In addition, data logging is possible via DATA-LOG Okolab software and via following third-party Imaging
software: LASX 2.0, NIS-Elements, SlideBook, MetaMorph and MicroManager. Macro for ZEN Black software is
available upon request.
UNO-STAGE-TOP-INCUBATOR iso also capable to receive TTL input (requires to order separately TTL-
CABLE).