47
a
b
Figure 67. Settings
–
Calibrations
–
Obj. Heater
–
Calibrate (a
–
b).
By clicking
Options
(see pointer 1 in Figure 68 a), the menu
Advanced calibration parameters
will
appear (see Figure 68 b) and, here, it is possible to set the calibration Accuracy (see pointer 2 in Figure
68 b) and Precision (see pointer 3 in Figure 68 b). Press
Default
(see pointer 4 in Figure 68 b) to
recover factory settings and
OK
(see pointer 4 in Figure 68 b) to confirm.
a
b
1
2
3
4
5
Figure 68. Settings
–
Calibrations
–
Obj. Heater
–
Calibrate
–
Options (a
–
b).
If you operate in
Sample Control
mode, you need to press
Calibrate
button and then make the Objective
Heater touch the sample in both oil immersion and water dipping imaging.
8.2.6
Logging
UNO-CONTROLLER is equipped with on-board memory for data logging and storage. A simple routine
allows downloading data to USB drive or to PC. In order to log and download the logged data, connect a USB drive
(not supplied) to UNO-CONTROLLER, using the supplied MINI-USB-OTG cable (see paragraph 7.6.1).
To activate the logging, follow the steps indicated below:
1.
Press on Settings icon
(see Figure 69 a) and press on Logging icon
(see Figure 69 b).