37
a
b
Figure 47. How to enter the Humidity menu (a -b).
GF-MIXER-HM (Humidity Module) supplies a humidity level up to90% of Relative Humidity acting on the
temperatures of both the Humidifier Bottle and the Heated Tube. In order to make the humidity control work
properly, the user has to set the actual Room Temperature- Press the
+
/
-
icons to set the Room Temperature (see
pointer 1 in Figure 48 b) and press
Save
to confirm (see pointer 2 in Figure 48 b).
a
b
1
2
Figure 48. Settings - Humidity
–
menu (a- b).
It is possible to act on the Humidity Module Offset if the desired humidity level cannot be achieved.
The humidity level of 90% of Relative Humidity may not be guaranteed if the Room Temperature
differs by more than ±1°C from 23°C and the GF-MIXER-HM (Humidity Module) is installed near
heating and air conditioning vents or near windows and doors that are frequently open.
By pressing the
Offset
button (see pointer 1 in Figure 49 a), the Offset subpage will be opened. Press the
+
/
-
icons to insert the Offset value (see pointer 2 in Figure 49 b) and press
OK
to save (see pointer 3 in Figure 49
b). Reduce the Offset value to increase the actual humidity level and, on the contrary, increase the Offset value to
lower the humidity content.