Figure 55. T Sensor attached on the bottom of the Petri dish (a -b).
Tip
►
Use the same dish that you use in regular practice.
•
Fill the sample holder with water or oil to the level used during standard operation. Make sure the tip
of the T Sensor is immersed in the liquid. Cover the sample holder with the lid or leave it open,
according to the protocol followed during standard operation.
•
Close the incubating chamber with its lid.
•
Press
Calibrate
button (see Figure 56 a) to start the Chamber calibration procedure (see Figure 56 b).
Click on
Abort
button to interrupt the procedure (see Figure 56 b).
a
b
Figure 56. Settings
–
Calibrations
–
Chamber
–
Calibrate (a
–
b).
By clicking
Options
(see pointer 1 in Figure 57 a), the menu
Advanced calibration parameters
will appear
(see Figure 57 b) and, here, it is possible to set the calibration Accuracy (see pointer 2 in Figure 57 b) and
Precision (see pointer 3 in Figure 57 b). Press
Default
(see pointer 4 in Figure 57 b) to recover factory
settings and
OK
(see pointer 4 in Figure 57 b) to confirm.