Defender 5000 Indicators Service Manual
1.2 SERVICE FACILITIES
To service an indicator, the service area should meet the following requirements:
Must be protected from electrostatic discharge.
Should be temperature controlled and meet the indicator specifications for temperature
environmental requirements. See specifications for temperature range.
Must be free of vibrations such as fork lift trucks close by, large motors, etc.
Must be free of air currents or drafts from air conditioning/heating ducts, open windows,
people walking by, fans, etc.
Area must be clean and air must not contain excessive dust particles.
Work surface must be stable and level.
Work surface must not be exposed to direct sunlight or radiating heat sources.
1.3 TOOLS AND TEST EQUIPMENT REQUIRED
In order to properly service the Indicator, certain special tools and test items are required in addition to
standard electronic tool kits. These items are listed as follows:
1.3.1 Special Tools and Test Equipment List
1. Ohaus Scale Base.
2. Load Cell Simulator optional.
3. Computer with RS232 Interface for testing the RS232 communications.
4. RS232 Interface cable.
5. Data Printer for use with RS232 communications.
6. ESD work station or mat.