18
Defender 5000 Indicators Service Manual
Power On Zero
Off,
On
Power On Unit
Auto
, kg, lb, g, oz, lb:oz, t, ton, c
Key Beep
Off,
On
Beep Volume
Low,
High
Transaction Counter
Off,
On
Next Transaction
1-9999999
Password
Off
, On
Set Password
xxxxxx
Confirm Password
xxxxxx
Reset
2.4.1 Capacity Unit
Select the unit used for calibration.
Kg
t (Metric Tonne)
lb
ton (Short Ton)
g
2.4.2 Range
Set the number of weighing ranges.
The TD52 terminals can be configured to use single or dual interval. Each interval can be assigned its own
graduation. If dual interval is selected, the graduation will change when the weight reaches the second
interval.
When
Single
range
is selected, the additional parameters available are:
>|1|< Capacity
>|1|< Graduation
When
Dual
range is selected, the terminal functions with two ranges, each with its own capacity and
graduation. In addition to the Range 1 capacity and graduation parameters, the following two parameters
are available:
>|2|< Capacity
>|2|< Graduation
2.4.3 Capacity
Set the scale capacity from 1 to 999999.
>|1|< Capacity
Specify the weight capacity for range 1. If
Single
range is enabled, this will be the scale capacity. If
Dual
range is enabled, this will be the first range.
>|2|< Capacity
Specify the weight capacity for range 2. If
Dual
range is enabled, this will be the scale capacity and it must
be bigger than >|1|< Capacity. If
Single
range is enabled, this parameter will not be shown.
2.4.4 Graduation
Set the scale readability from 0.0001 to 100.