2.7.8 Custom Unit (c)
Use the Custom Unit to display weight in an alternative unit of measure. The custom unit is defined using a
conversion factor, where the conversion factor is the number of custom units per gram expressed in
scientific notation (Factor x 10^Exponent).
Factor
Set the conversion factor using the numeric keypad.
Settings of 0.1000000 to 1.9999999 are available. The default setting is 1.0.
Exponent
Set the factor multiplier.
-3 = divide the Factor by 1000 (1x10
-3
)
-2 = divide the Factor by 100 (1x10
-2
)
-1 = divide the Factor by 10 (1x10
-1
)
0
= multiply the Factor by 1 (1x10
0
)
1 = multiply the Factor by 10 (1x10
1
)
2 = multiply the Factor by 100 (1x10
2
)
Least Significant Digit(LSD)
Set the graduation.
Settings of 0.5, 1, 2, 5, 10, 100 are available.
The Custom Unit’s name can be customized up to 3 characters.
Note:
Custom Unit is locked at Off position when the Security Switch is set to the locked position.
Custom Unit is not available when Range is set to Dual interval.
Set the status.
Off
= Disabled
On = Enabled