Defender 5000 Indicators Service Manual
Clear the pan and then press the soft key
corresponding to the icon
.
The message 'Completed' will be displayed on the
screen.
Exit Zero Calibration by pressing the soft key
corresponding to the icon
.
To return to the Main Menu, press the soft key
corresponding to the icon
.
2.3.2 Span Calibration
Span Calibration uses one point. The span calibration point is established with a calibration mass placed
on the scale.
Note:
Span Calibration should be performed after Zero Calibration.
Calibration procedures:
Long press the button
to enter the Main
Menu.
Press the soft key corresponding to the icon
to
enter the Calibration sub-menu.
Scroll to Span Calibration using the soft key
corresponding to the icon
.
Press the soft key corresponding to the icon
to
initiate Span Calibration.
Place a calibration mass of the specified weight on
the pan and press the soft key corresponding to the
icon
. To change to a different calibration point,
input the value desired, and then place the
corresponding weight on the pan for calibration.
A suggestive message shows on the screen.
The message 'Completed' will be displayed on the
screen.