Defender 5000 Indicators Service Manual
Exit Span Calibration by pressing the soft key
corresponding to the icon
.
To return to the Main Menu, press the soft key
corresponding to the icon
.
Note: Span Calibration should be performed after
Zero Calibration.
2.3.3 Linearity Calibration
Linearity calibration uses 3 calibration points. The full calibration point is established with a weight on the
scale. The mid calibration point is established with a weight equal to half of the full calibration weight on
the scale. The zero calibration point is established with no weight on the scale. The mid calibration points
cannot be altered by the user during the calibration procedure.
Calibration procedures:
Long press the button
to enter the Main
Menu.
Press the soft key corresponding to the icon
to
enter the Calibration sub-menu.
Scroll to Linearity Calibration using the soft key
corresponding to the icon
.
Press the soft key corresponding to the icon
to
initiate Linearity Calibration.
Clear the pan and press the soft key corresponding
to the icon
.
Put the calibration mass of the specified weight on
the pan, and then press the soft key corresponding
to the icon
for confirmation. To change to a
different calibration point, input the value desired,
and then place the corresponding weight on the
pan for calibration.
Put the calibration mass on the pan, and then
press the soft key corresponding to the icon
for
confirmation. To change to a different calibration
point, input the value desired, and then place the
corresponding weight on the pan for calibration.
The message 'Completed' will be displayed on the
screen.