20
Defender 5000 Indicators Service Manual
2.4.9.1 Next Transaction
The value of the next transaction displays in the Next Transaction field.
2.4.10 Password
Set the password.
Off
= the menu is accessed freely without password.
On = the menu is accessed only after entering a password up to 6 digits.
2.4.11 Reset
Reset the Setup menu to the factory defaults (except Range, Capacity and Graduation).
No
= not reset.
Yes = reset.
NOTE:
If the Security Switch is set to ON, the Capacity Unit, Range, Capacity, Graduation and Power On
Zero settings are not reset.
2.5 Readout Menu
Enter this menu to customize display functionality. Default settings are
bold
.
Read Out
Options
Stability
0.5d,
1d
, 2d, 5d
Zero Range
+/-2%, +/-
100%
Filter Level
Low,
Medium
, High
Auto Zero Track
Off,
0.5d
, 1d, 3d
Backlight
Off,
1min
, 2min, 5min,10min, Always On
Screensaver
Off,
5min
, 10min, 30min
Auto Off
Off
, 5min, 10min, 30min
Base Auto Off
Off, 5min,
10min
, 30min
Adjust Contrast
1, 2,
3
, 4, 5
Reset
2.5.1 Stability
Set the amount the reading can vary before the stability symbol turns off.
0.5d = 0.5 scale division
1d
= 1 scale division
2d = 2 scale divisions
5d = 5 scale divisions
2.5.2 Zero Range
Set the percentage of scale capacity that may be zeroed.
2%
100%
NOTE:
The setting is forced and locked to 2% when the Security Switch is set to the locked position.
2.5.3 Filter Level
Set the amount of signal filtering.
Low
= faster stabilization time with less stability.