![Oerlikon CENTER TWO Operating Manual Download Page 69](http://html1.mh-extra.com/html/oerlikon/center-two/center-two_operating-manual_3398549069.webp)
68
GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
Computer interface
6.3.10
EUM
Switch the emission. See Chapter 5.2.10 Emission (EMi),
45.
S: EUM[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
6.3.11
FIL
Filter. Measurement filter. See Chapter 5.2.1 Measurement filter (FiLt),
S: FIL[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
6.3.12
FSR
Full scale range. Full scale range of linear transmitters (CTR, DI). See
Chapter 5.2.3 Measuring range (FS),
42.
S: FSR[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
Parameters
Significance
a
Emission for channel 1
0 = Manually
1 = Automatic (default)
b
Emission for channel 2
(see above)
c
Emission for channel 3
(see above)
Parameters
Significance
a
Filter for channel 1
0 = Fast
1 = Medium (default)
2 = Slow
3 = CTR
b
Filter for channel 2
(see above)
c
Filter for channel 3
(see above)