Parameters
5.2.3
Measuring range (
FS
)
Linear transmitters (CTR) require specification of the full-scale range. You can
set this value using the cursor buttons. The following values are available:
0.01 mbar
0.01 Torr, 0.02 Torr, 0.05 Torr
0.10 mbar, 0.25 mbar, 0.50 mbar
0.10 Torr, 0.25 Torr, 0.50 Torr
1 mbar, 2 mbar, 5 mbar
1 Torr, 2 Torr, 5 Torr
10 mbar, 20 mbar, 50 mbar
10 Torr, 20 Torr, 50 Torr
100 mbar, 200 mbar, 500 mbar
100 Torr, 200 Torr, 500 Torr
1000 mbar, 1100 mbar
1000 Torr
2 bar, 5 bar, 10 bar, 50 bar
DI gauge heads
200 mbar (
di200
)
2 bar (
di2
)
2 bar relative (
dir2
)
5.2.4
Offset (
oFS
)
When the offset correction is activated, a previously specified offset value will
be subtracted from each pressure reading. This allows to conveniently mea-
sure the relative pressure with respect to a reference pressure.
The offset correction affects the readings on the display, the RS232C output,
the recorder output, and the switching functions. The analog outputs, how-
ever, are not affected.
Establishing an offset value and activating the offset correction
1
Select the
oFS
parameter
2
Keep the UP button pressed for approximately 2 seconds
The current pressure reading becomes the new offset value
The offset correction is activated
The OFS indicator is illuminated
Deactivating the offset correction
1
Select the
oFS
parameter
2
Press the DOWN button
The offset correction is deactivated
The display shows
oFF
The OFS indicator is dark
Activating the offset correction
1
Select the
oFS
parameter
2
Press the UP button