52
GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
Parameters
5.4
Test parameters (
PArA tESt
)
This parameter group is intended for test and service purposes. It is used to
examine additional system data, to set basic system parameters, and to check
individual system functions.
The parameter group
tESt
is not required for normal operation. For this reason
it is not accessible normally.
5.4.1
Selection
When switching on
The parameter group
tESt
becomes available if you switch on the Multi-Chan-
nel Controller as follows:
1
Press the PARA button and keep it pressed
2
Switch the mains switch on. See Fig. 3-4,
18, item H.
The Multi-Channel Controller is switched on
The parameter group
tESt
is selected
The PARA indicator is illuminated
During normal operation
During normal operation it is also possible to activate the parameter group
tESt
from the measurement mode:
1
Press the PARA button
The parameter menu is selected
The PARA indicator is illuminated
2
Keep the UP and DOWN buttons pressed for approximately 5 seconds
The firmware version is displayed
The parameter group
tESt
is selected
Note:
When the parameter group
tESt
is selected, the Multi-Channel Controller
will not automatically return to the measurement mode. In order to return
to the measurement mode, press the PARA button repeatedly until all
test parameters have been run through.
5.4.2
Firmware version (
Pnr
)
Displays the firmware version number. The last character represents the mod-
ification index.
Example: 302-533-F