GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
Computer interface
6.3.6
DCD
Display control digits. Number of digits shown in the display. See
Chapter 5.3.3 Display format (diGit),
48.
S: DCD[,a]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a<CR><LF>
Note:
When
PrE
is enabled, the display of TTR transmitters in the pressure
range p < 10
-4
mbar is reduced by one digit.
6.3.7
DGS
Degas. See Chapter 5.2.5 Degas function (dEGAS),
43.
S: DGS[,a,b,c]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
Note:
The degas function is switched off automatically after 3 minutes. It may
be also be stopped prematurely.
Parameters
Significance
a
Number of digits
2 = 2 digits (default)
3 = 3 digits
Parameters
Significance
a
Transmitter 1
0 = Degassing off (default)
1 = Degassing on
b
Transmitter 2
(see above)