5: How the Flame Spectrometer Works
225-00000-000-11-201604
41
Grating
Number
Intended Use Groove Density Spectral Range Blaze Wavelength
Best Efficiency
(>30%)
UV-VIS
2400
50-120 nm
Holographic VIS
260-780 nm
NIR
600
625 nm
1000 nm
650-1100 nm
UV-NIR
500
200-1025 nm
250 nm
200-450 nm
Grating efficiency curves are available to view by using our online range and resolution
calculator at
www.oceanoptics.com/product-category/modular-
Gratings Showing Light Diffracted into its Constituent Wavelengths
6.
Focusing Mirror (specify standard or SAG+):
This mirror focuses first-order spectra
on the detector plane. Both the collimating and focusing mirrors are made in-house to
guarantee the highest reflectance and the lowest stray light possible. You can opt to
install a standard or SAG+ mirror. As with the collimating mirror, the mirror type needs
to be specified when ordering.
7.
Detector Collection Lens (optional):
This cylindrical lens is fixed to the detector to
focus the light from the tall slit onto the shorter detector elements. It increases light-
collection efficiency and reduces stray light. It also is useful in a configuration with a
large-diameter fiber and slit for low light-level applications such as fluorescence.
Preconfigured Flame spectrometers with a collector lens are available – look for –ES at
the end of the name.
8.
Detector:
There are two choices of detector available for the flame. We offer a 2048-
element FLAME-S (Sony ILX511B) or a 3648 element FLAME-T (Toshiba
TCD1304AP) linear CCD array. These both have an effective range of 190-1100 nm. The
optics split the light into its component wavelengths which fall across the different
Summary of Contents for Flame-S
Page 18: ...1 Introduction 8 225 00000 000 11 201604...
Page 32: ...2 Installation and Setup 22 225 00000 000 11 201604...
Page 46: ...4 Troubleshooting 36 225 00000 000 11 201604...
Page 54: ...5 How the Flame Spectrometer Works 44 225 00000 000 11 201604...
Page 74: ...7 Calibration 64 225 00000 000 11 201604...
Page 130: ...Index 120 225 00000 000 11 201604...