![Ocean Optics Flame-S User Manual Download Page 117](http://html1.mh-extra.com/html/ocean-optics/flame-s/flame-s_user-manual_3228091117.webp)
8: Firmware and Advanced Communications
225-00000-000-11-201604
107
Letter
Description
Version
E
F
G
Set Data Compression
1.00.0
H
I
Sets integration time (ms increments)
1.00.0
J
Sets Lamp Enable Signal
1.00.0
K
Changes baud rate
1.00.0
L
Clear Memory
M
N
O
P
Partial Pixel Mode
1.00.0
Q
R
S
Starts spectral acquisition with previously set parameters
1.00.0
T
Sets trigger mode
1.00.0
U
V
W
Set FPGA Register Information
1.00.0
X
Y
Z
Read out Scan from memory
1.00.0
A
Set ASCII mode for data values
1.00.0
b
Set binary mode for data values
1.00.0
i
Set integration value (32-bit value and us increments)
1.00.0
k
Sets Checksum mode
1.00.0
v
Provides microcode version #
1.00.0
x
Sets calibration coefficients
1.00.0
?
Queries parameter values
1.00.0
+
Reads the plugged-in accessories
1.00.0