5: How the Flame Spectrometer Works
40
225-00000-000-11-201604
Reflectance vs. Wavelength for Aluminum, Gold, and Silver Mirrors
By Bob Mellish in Wikipedia
5.
Grating and Wavelength Range (specify grating and starting wavelength):
We
install the grating on a platform that we then rotate to select the starting wavelength
you have specified. Then we permanently fix the grating in place to eliminate
mechanical shifts or drift.
Use our online
Range and Resolution Calculator
to find out how your grating choice
affects spectral range and optical resolution by viewing the grating efficiency curves.
The available gratings are summarized below:
Grating
Number
Intended Use Groove Density Spectral Range Blaze Wavelength
Best Efficiency
(>30%)
UV
600
650 nm
300 nm
200-575 nm
UV-VIS
600
650 nm
400 nm
250-800 nm
VIS-Color
600
650 nm
500 nm
350-850 nm
NIR
600
625 nm
750 nm
530-1100 nm
UV-VIS
1200
300 nm
Holographic UV
200-400 nm
NIR
1200
200-270 nm
750 nm
500-1100 nm
UV-VIS
2400
100-140 nm
Holographic UV
200-500 nm
VIS-NIR
1200
200-270 nm
Holographic VIS
400-800 nm
UV-VIS
1800
100-190 nm
Holographic UV
200-635 nm
UV-VIS
1800
120-160 nm
Holographic VIS
320-720 nm
Summary of Contents for Flame-S
Page 18: ...1 Introduction 8 225 00000 000 11 201604...
Page 32: ...2 Installation and Setup 22 225 00000 000 11 201604...
Page 46: ...4 Troubleshooting 36 225 00000 000 11 201604...
Page 54: ...5 How the Flame Spectrometer Works 44 225 00000 000 11 201604...
Page 74: ...7 Calibration 64 225 00000 000 11 201604...
Page 130: ...Index 120 225 00000 000 11 201604...