![Ocean Optics Flame-S User Manual Download Page 115](http://html1.mh-extra.com/html/ocean-optics/flame-s/flame-s_user-manual_3228091115.webp)
8: Firmware and Advanced Communications
225-00000-000-11-201604
105
Query Status
Returns a packet of information containing the current operating information. The structure of
the status packet is given below:
Byte Format
Byte 0
0xFE
Return Format
The data is returned in Binary format and read in by the host through End Point 1 In. The
structure for the return information is as follows:
Byte
Description
Comments
0-1
Number of Pixels - WORD
LSB | MSB order
2-5
Integration Time - WORD
Integration time in
s
– LSW | MSW.
Within each word order is LSB | MSB
6
Lamp Enable
0
– Signal LOW
1
– Signal HIGH
7
Trigger Mode Value
8
Spectral Acquisition Status
9
Packets In Spectra
Returns the number of Packets in a
Request Spectra Command.
10
Power Down Flag
0
– Circuit is powered down
1
– Circuit is powered up
11
Packet Count
Number of packets that have been
loaded into End Point Memory
12
Reserved
13
Reserved
14
USB Communications Speed
0
– Full Speed (12Mbs)
0x80
– High Speed (480 Mbps)
15
Reserved
Summary of Contents for Flame-S
Page 18: ...1 Introduction 8 225 00000 000 11 201604...
Page 32: ...2 Installation and Setup 22 225 00000 000 11 201604...
Page 46: ...4 Troubleshooting 36 225 00000 000 11 201604...
Page 54: ...5 How the Flame Spectrometer Works 44 225 00000 000 11 201604...
Page 74: ...7 Calibration 64 225 00000 000 11 201604...
Page 130: ...Index 120 225 00000 000 11 201604...